QUI-1122
Page 39
Markes International Ltd
T:
+44 (0)1443 230935
F:
+44 (0)1443 231531
www.markes.com
6.3
UNITY–Air Server-xr Electronic Gas Control with the Shimadzu QP 2010 (Plus) GCMS fitted with
AFC
NOTE
: To complete this installation the specialised GM - 1/16” fittings must be ordered directly from
Shimadzu prior to the installation. The parts required are as follows:
Part no.
Description
Quantity
221-35003
Shimadzu fitting GM-1.6L
1
221-35004-91
Shimadzu fitting GM-1.6L
1
228-16000-84
Ferrules
4
228-16002-84
Nuts
2
6.3.1
Gas connections between UNITY–Air Server-xr and the Shimadzu AFC
The Shimadzu GC is fitted with two molecular sieve filters, one in the carrier gas line and one in the
septum purge line. Remove the filter from the carrier gas line and replace with the Shimadzu fitting
GM-1.6L. The filter on the septum purge line should remain in place to prevent back diffusion of
air/water.
UNITY-xr