Ensure that there is a pre-trap fire purge time of at least 1.0 min and that the “Trap High (C)” is compatible
with the temperatures given in the Trap Certificate. If any of the default settings are changed the software
will prompt you to save it with a new method name.
Click the Sequence tile on the MIC home screen, and insert a blank sample using the Auto assign function
to select the appropriate Trap conditioning method (either the default method or a modified one if the
conditions required are different from the default).
NOTE:
The tube number entered into the sequence becomes irrelevant when running a tube
conditioning method.