![Markes International UNITY-Air Server-xr Скачать руководство пользователя страница 18](http://html.mh-extra.com/html/markes-international/unity-air-server-xr/unity-air-server-xr_installation-manual_1727984018.webp)
QUI-1122
Page 18
Markes International Ltd
T:
+44 (0)1443 230935
F:
+44 (0)1443 231531
www.markes.com
Remove the back cover by sliding it backwards.
Place the 1/16 x 0.4 mm (U-FV001 (pk 10)) ferrule
onto the union then thread the 1/16-inch stainless
steel Swagelok type nut (SERZ-0157) on loosely
Pull ~ 20 cm of fused silica from the PTFE line
casing. Cut off the first few mm of fused silica
using an approved capillary cutting tool.
Make a mark 20 mm from the end of the fused
silica using typing white-out fluid or an alternative
marker.
Gently feed the fused silica through the 1/16” nut
and ferrule until the mark is reached.
Tighten the nut to trap the fused silica and then
tighten a further 1/2 turn using one of the 8 mm
wrenches (spanners) provided in the shipping kit.
Do not over tighten or the ferrule will
become distorted.
20 mm
insertion
depth