QUI-1122
Page 35
Markes International Ltd
T:
+44 (0)1443 230935
F:
+44 (0)1443 231531
www.markes.com
6.
Configuration of UNITY–Air Server-xr Electronic Gas Control – instrument specific
guides
6.1
UNITY–Air Server-xr Electronic Gas Control with the Agilent 6890GC
NOTE:
To complete this installation the following 6890 parts are recommended. They can be ordered
directly from Agilent.
Part no.
Description
Quantity
G2131-80500
Replacement 2 port EPC block
1
1390-1022
Screw
1
However, if these parts are not available at installation both the carrier gas and septum purge lines
running between the EPC module and the Split/Splitless injector must be cut and extended using 1/8”
unions and green PEEK tubing in order to facilitate connection to UNITY–Air Server-xr.
NOTE:
When installing onto an existing 6890GC the firmware on the GC must be A.03.08 or later (for A-
series) or N.04.09 (for N-series). To check the firmware version, use the keyboard on the GC,
press > Options > Diagnostics > Instrument Status and then scroll down to ‘Version’ where you will
see the version of the firmware running on the instrument.
6.1.1
Procedure
Remove the top rear cover on the GC to provide access to the EPC module that is going to be used to
regulate the carrier gas supply to UNITY–Air Server-xr.
Removing the rear panel to access EPC modules
Remove the 3 port EPC block assembly from the split/splitless EPC module.
Fit the replacement EPC block assembly, and route the 1/16” tubing through the rear panel of the GC
and towards the UNITY–Air Server-xr.