2.4.3
Pressure controlled carrier gas supply
Helium is invariably used as the carrier gas for capillary chromatography and nitrogen for packed column
or sensor work. 5.0 grade (i.e. 99.999%) or higher purity helium / nitrogen gas is recommended in either
case.
UNITY–Air Server-xr requires a regulated supply of carrier gas at a pressure to suit the analytical column /
system selected. The UNITY–Air Server-xr gas flow path has minimum (<2 psi) impact on total system
impedance. Suitable pneumatic control for the carrier gas may already be available on your GC. The
performance of most common capillary columns is optimised at between 1 and 2 ml/min typically
requiring between 10 and 30 psi head pressure.
Previously approved and correctly rated external gas regulator should be used with this unit. Maximum
inlet pressure must not exceed 60 psi.
The pressure in the laboratory carrier gas line should be at least 10 psi higher than that supplied to
UNITY–Air Server-xr.
2.4.4
Filters
Deoxo and organic filters should be included in the carrier gas line just upstream of connection to the
UNITY–Air Server-xr - GC analytical system (see Accessories for Thermal Desorption catalogue).