![Markes International UNITY-Air Server-xr Скачать руководство пользователя страница 14](http://html.mh-extra.com/html/markes-international/unity-air-server-xr/unity-air-server-xr_installation-manual_1727984014.webp)
QUI-1122
Page 14
Markes International Ltd
T:
+44 (0)1443 230935
F:
+44 (0)1443 231531
www.markes.com
3.2.4
Cable connections
Connect the power cable (
SERZ-0024
).
Connect the UNITY–Air Server-xr to the PC using the serial cable supplied in the shipping kit (
SERZ-0189
).
If there are no available serial ports on the PC, use the USB-Serial cable (
U-USBSR
) provided.
Connect the GC cable. Depending upon the GC that is being used a different connection cable will be
required:
SERUTE-5142
= AGILENT GC
SERUTE-5143
= THERMO GC
SERUTE-5144
= VARIAN GC
SERUTE-5141
= SHIMADZU / OTHER GC
3.3
Installing the Transfer Line
UNITY–Air Server-xr is supplied with a universal transfer line (
SERUTE-5330-xr
) to convey desorbed
analytes from UNITY–Air Server-xr to a gas chromatograph or other analytical system. The sample path
utilises a deactivated fused silica line (0.25 mm I.D. an d 0.35 mm O.D.) heated over its entire length by
means of a distributed heater and at the GC end by heat conduction from the GC oven. The line is 1.4 m
long, which is sufficient to reach most gas chromatographs even when a mass spectrometer is attached.
NOTE
: The parts supplied can be used in other combinations to suit particular instrument configurations.
3.3.1
Installing the Transfer Line onto the GC
There are two options for transfer line installation; one requires use of a hole in the top, side or back of the
GC oven, while the other option is to gain access via an unused heated injector port.
Option 1 – Through Hole in Oven Wall
1.
Locate a hole in the inner oven wall with a corresponding hole leading to the outside of the
instrument.
2.
Where necessary, displace the oven insulation material such that the transfer line heater can be
fitted directly against the external face of the oven wall.
3.
Place one of the washers provided in the UNITY-xr shipping kit on the external face of the oven
wall, over the hole through which the transfer line will be passed.
NOTE
: A number of different washers are provided. The washer required will be determined by the
hole in the oven through which the transfer line will pass.
4.
Fit the GC end of the UNITY-xr transfer line through the hole in the washer and on through the hole
in the GC oven wall.