![Markes International UNITY-Air Server-xr Скачать руководство пользователя страница 40](http://html.mh-extra.com/html/markes-international/unity-air-server-xr/unity-air-server-xr_installation-manual_1727984040.webp)
QUI-1122
Page 40
Markes International Ltd
T:
+44 (0)1443 230935
F:
+44 (0)1443 231531
www.markes.com
The carrier gas out line from the Shimadzu AFC (“CG Out”) is now connected to the “Carrier Inlet” port
on the back of UNITY–Air Server-xr using a 1 m length of green PEEK tubing.
The “Pressure Outlet” from the back of UNITY-xr is now connected - with a similar length of green
PEEK tubing - to the molecular sieve filter on the septum purge line (and hence to the (“SP In”)) on
the Shimadzu AFC. It is important that the two lengths of green PEEK tubing are the same length.
Carrier gas is supplied and regulated upstream of the system in the supply line and it’s pressure is
monitored downstream of the system in the septum purge line.
6.3.2
Configuring the Shimadzu GC software
The inlet Split Mode on the Shimadzu GC must be set to Direct.
a.
To access the Split Mode press the “FLOW” key on the GC keypad.
NOTE:
In the case of GCs with two or more AFC units it may be necessary to press the “FLOW” key
repeatedly until the correct one is displayed.
b.
Move the flashing cursor over the “Split Mode” (type using the arrow keys on the GC keypad).
c.
Using the right arrow key, select “DIRECT” mode as shown, and press the “ENTER” key on the GC
keypad.
Setting Inlet Mode as Direct on Shimadzu GC
Shimadzu fitting on
carrier gas line
Molecular Sieve filter on
septum purge line
Two lengths of PEEK tubing
connecting to UNITY-xr