Once the current configuration section is correct the communications port section will update and allow
the default communications ports to be changed if necessary by clicking on the drop down box and then
clicking on the correct Com port in the list. The correct communication port designations can be located in
Control Panel > Device manager > Ports
NOTE
: the correct ports will depend on the PC setup and whether the supplied serial-USB adapters are
required.
The UNITY–Air Server-xr is shipped with two MFCs, one MFC controls split and dry purge (in tube
desorption mode only) and one is used to control the trap flow. To set them and carrier gas options up,
click onto the Gas and Flows Tab.
Any MFC’s detected will be shown, along with their address and available gases. To designate the MFCs,
click on the drop down box for the MFC using Address 1 and select split and for the MFC using Address 2
select trap.
Select the appropriate carrier gas, e.g Helium, from those available in the carrier gas drop down box.