Instruction Manual
MAN-001
38
www.macmedical.com
Steel Door Hinge Reversal and Re-installation
1. Loosen the nuts on the bolts that hold the top door hinge onto the
unit. Hold the door as you remove the nuts so the door does not fall
on you. Once the nuts are removed, the door can be leaned away
from the unit and lifted off the bottom hinge (Fig. 58).
2. Remove the nuts that fasten the bottom hinge to the unit and remove
the screws and bottom hinge (Fig. 59).
note:
For warming cabinets with more than one door, an
intermediate hinge (Fig. 61) fits between the bottom of the top
door and the top of the bottom door of the dual chambered
warming cabinet (shown in blue in Fig. 62. (Middle doors on Triple
chamber warming cabinets use only intermediate hinges for both
top and bottom.)
Intermediate hinges come as either right hand (part #W0015) or left hand (part #W0016) hinges. Purchase the
intermediate hinge that is the opposite of the current one on your warming cabinet.
If the unit has multiple doors, remove the intermediate hinges by unscrewing them from the cabinet.
Purchase Parts needed for Door Hinge Reversal
•
Intermediate Hinge (for multiple door units only) W0015 (Right Hand)
or W0016 (Left Hand). Obtain the intermediate hinge opposite of the
currently installed hinge.
W0016
Left hand
W0015
Right hand
Fig. 58: Remove Steel Door
Fig. 59: Remove Bottom Hinge
Fig. 60: Steel Door Hinge Parts
Fig. 61: Intermediate Door Hinge
Fig. 62: Intermediate Door Hinge in Place
Содержание D Series
Страница 41: ...Instruction Manual MAN 001 41 www macmedical com Wiring Diagram Single Chamber Warmers ...
Страница 42: ...Instruction Manual MAN 001 42 www macmedical com Wiring Diagram Dual Chamber Warmers ...
Страница 43: ...Instruction Manual MAN 001 43 www macmedical com Wiring Diagram Triple Chamber Warmers ...