Glass Door Hinge Reversal and Re-installation
1. Using a drift pin (or a small headed screwdriver) and a
hammer, detach the glass door from the unit by tapping
lightly on the hinge pins, driving them through the two
halves of the hinges. Then pull the door away from the
unit to remove it (Fig. 45).
2. Detach the male halves of the glass door hinge by
unscrewing them from the warming cabinet. The hinge
halves on the door should remain in place (Fig. 46).
Fig. 44: Glass Door Hinge Parts
Fig. 45: Remove Door Hinge Pins
Fig. 46: Detach Glass Door Hinge - Male Half
Following are instructions for the removal, hinge
reversal, and re-installation of Glass Doors.
note:
No additional parts are required to perform
Glass Door hinge reversal. However, it may be nec-
essary to drill holes in the bottom of the Glass Door
to enable relocation of the Cam Lock Plate to its
new position. The following tools will be necessary
for this procedure:
•
1/8” Drill Bit
•
8/32 Tap