Instruction Manual
MAN-001
34
www.macmedical.com
Fig. 47: Remove Cam Lock and Handle
3. Unscrew the Door Handle and Cam Lock Plate from their present position and move them to the other
side of the door. Make sure to “mirror” the Cam Lock Plate so that its top flange will be flush against the
warming cabinet when the door is re-installed in its new position (Fig. 47).
4. If it is necessary to drill holes for the relocation of the Cam Lock Plate on the other side of the door,use the
Cam Lock plate as a guide to mark the proper location of the holes to be drilled (Figure 48). The Cam Lock
Plate should be positioned 1/8” from the outer edge of the door (Figure 49).
Fig. 48: Use Cam Lock for hole guide
Fig. 49: Proper position of Cam Lock
Cam Lock Plate set back
1/8” from outer face of door
Door Inner Face