Installing D-Series Warming Cabinets
environmental Conditions
This unit is intended for use in a stable ambient
environment, with an ideal temperature of 72° F
(22.22° C) or less. The unit should never be used
directly next to any appliance that may produce
heat, such as an autoclave.
During Transport and Storage (in original packaging
materials) -
•
Ambient Temperature: -40° - 159°F (-40° - 70°C)
•
Relative Air Pressure: 10% - 100%, including
condensation
•
Air Pressure: 500 hPa (14 inHg - 31.3 inHg)
During Use - for Dry Locations
•
Ambient Temperature: 60° - 85°F (15° - 30°C)
•
Relative Air Moisture: 30% - 60% non-condensing
•
Air Pressure: 700 hPa - 1060 hPa (20.7 inHg - 31.3
inHg)
Installation
Before starting the installation, review the local
electric codes including the Occupational Health
and Safety Act for any requirements pertaining to the
proper installation of this equipment.
Contact your MAC Medical representative for seismic
calculations and tie-down hardware, if applicable.
1. Carefully uncrate the D-Series Warming Cabinet.
2. Inspect for any damage. If there is damage,
please contact MAC Medical, Inc. at (1-877-828-
9975).
3. Check your 120V, 60 Hz, Single Phase 15 AMP
GFIC Protected electrical outlet or 220V, 60 Hz,
Single Phase, 7 AMP GFIC Protected electrical
outlet. Be sure the outlet is safely accessible and
in proper working condition.
4. Plug the 3-prong electrical plug into the 120VAC,
60Hz, 15 AMP, GFIC Protected Outlet or a 220
VAC, 60Hz, Single Phase, 7 AMP GFIC protected
electrical outlet. Make sure the electrical outlet is
safely accessible and in proper working condition.
5. Place Warming Cabinet on a solid, level platform
where external movement will not interfere with
loose contents used by the warmer.
6. Make sure the shelving is correctly located as
desired and level. If not, adjust their height (see
“Adjusting the Shelves” on page 19).
7. Before use, remove any items that have been
stored in the cabinet.
Testing before Using
All warming cabinets have been calibrated and
tested before leaving the factory. There is no need
for the user to do additional testing after installation
prior to use.
After six month of use, it is recommended that the
user test the warming cabinet for temperature
accuracy. See “Semi-Annual Checklist” on page
24.