90
47DAC OPERATING MANUAL EM-9518-B Rev.4
8.3 ANALOG OUTPUT 0% / ANALOG OUTPUT 100%
8.3.1 OPERATING PROCEDURE
NOTE
The following figures are display examples. The displays
depend on the specifications and settings.
1
Hold down Scale/
↑
button for 3 seconds or more to move on to
Scaling Setting Mode.
• The input type is indicated.
• The SD indicates ‘INTYPE’.
2
Press Alarm/
↓
or Scale/
↑
button to go to the analog output 0%
setting.
• The analog output 0% is indicated.
• The SD indicates ‘ANG L’.
• ‘Zro’ indicator turns on.
3
Press Shift button to shift the display into the setting standby
mode. Then press Shift and Up buttons to set the analog out-
put 0%.
• The sixth digit starts blinking, to which you can apply changes.
• Set within the range of -20000 to 100000.
NOTE
Set the analog output 0% with the decimal point position set in the decimal point position setting.
4
Press Alarm/
↓
button to apply the new value and go to the
analog output 100% setting.
• The analog output 0% is registered.
• The analog output 100% is indicated.
• The SD indicates ‘ANG H’.
• ‘Zro’ indicator turns off and ‘Spn’ turns on.
2,4,
6,7
1,2,
6,7
3,5
3,5
Blinking