47DAC OPERATING MANUAL EM-9518-B Rev.4
33.1.2 OPERATING PROCEDURE
NOTE
The following figures are display examples. The displays
depend on the specifications and settings.
1
Hold down Scale/
↑
button for 3 seconds or more to move on to
Scaling Setting Mode.
• The input type is indicated.
• The SD indicates ‘INTYPE’.
IMPORTANT
Warm up the unit for 10 minutes or more before carrying out the Teach Calibration.
2
Press Alarm/
↓
or Scale/
↑
button to go to the input scaling
value Zero setting.
• The input scaling value Zero is indicated.
• The SD indicates ‘IN-A’.
• ‘Zro’ and ‘Tch’ indicators turn on.
NOTE
Skip to Step 5 when the teach calibration (Zero) is not necessary.
3
Press Up button to go to the teach calibration (Zero) setting.
• The present input is indicated.
• ‘Tch’ indicator starts blinking in red.
NOTE
The decimal point position depends on the decimal point position setting. Disregard the decimal point to calibrate.
2,5,
8
1,2,
8
3,4,
6,7
Blinking