47DAC OPERATING MANUAL EM-9518-B Rev.4
13.2 LOW-END CUTOUT VALUE
13.2.1 OPERATING PROCEDURE
NOTE
The following figures are display examples. The displays
depend on the specifications and settings.
1
Hold down Alarm/
↓
and Scale/
↑
buttons at once for 3 seconds
or more to move on to Advanced Setting Mode.
• The averaging type is indicated.
• The SD indicates ‘AVE-TP’.
NOTE
With the I/O option code ‘6’ or ‘A’, the event trigger mode is indicated (‘EVENT’ on the SD).
2
Press Alarm/
↓
or Scale/
↑
button to go to the low-end cutout
value setting.
• The low-end cutout value is indicated.
• The SD indicates ‘ZLMTN’.
3
Press Shift button to shift the display into the setting standby
mode. Then press Shift and Up buttons to set the low-end
cutout value.
• The third digit starts blinking, to which you can apply changes.
• Set within the range of 000 to 999.
NOTE
Set the value for the display scaling. The decimal point is not indicated.