120
47DAC OPERATING MANUAL EM-9518-B Rev.4
10.3 BANK NO.
10.3.1 OPERATING PROCEDURE
NOTE
The following figures are display examples. The displays
depend on the settings.
1
Hold down Alarm/
↓
button for 3 seconds or more to move on
to Alarm Setting Mode.
• The bank No. is indicated.
• The SD indicates ‘ALM BNK’.
2
Press Shift and Up buttons select the bank No.
• Select one among ‘1’ through ‘8’.
3
Press Alarm/
↓
or Scale/
↑
button to apply the bank No.
• And the next parameter setting is indicated.
NOTE
• Press Alarm/
↓
button, and the alarm output pattern will be indicated (‘ALM PTN’ on the SD).
• Press Scale/
↑
button, and the bank copy will be indicated (‘BNK CPY’ on the SD).
4
■
TO GO ON TO SET THE ALARM SETPOINT,
Skip to Step 2 in “9.2 ALARM SETPOINT”.
■
TO QUIT,
Hold down Alarm/
↓
or Scale/
↑
button for 1 second or more to return to Measuring Mode.
1,3
3
2
2
Blinking