Chapter 4 Preparing LogTag(s) for use
41
Function
Purpose
Clear and reset alarm
when STOP/CLEAR
button pressed
If this option is enabled, any user can turn off an active alarm by
pressing the LogTag's STOP/CLEAR button. It will also reset the
internal counters such that it would be as if no alarm had occurred
while the LogTag was recording. Should the LogTag subsequently
detect a further alarm condition, the alarm will activate again. Once
the LogTag has finished recording, pressing the STOP/CLEAR
button will not clear and reset any active alarm.
In any case, an alarm will be registered in the day summary,
regardless whether it was cleared or not.
Alarm remains on even
if readings return to
non-violation range
If this option is enabled then the alarm symbol will stay on, even if
the current reading does not meet the alarm conditions specified by
the previous controls.
If this option is disabled then the alarm symbol will stay on only if
the current reading meets the alarm conditions specified by the
previous controls.
Pause alarm/statistics
processing for X
readings
If this option is enabled, pressing either button on the LogTag
while recording excludes the next reading from the alarm and
statistics calculations; PAUSED will now be shown on the display.
This allows the user to review the statistics or clear an alarm
without causing an invalid reading, alarm or statistic.
Switch off display after
30 seconds
If this option is enabled, the display will turn off after 30 seconds to
save battery power.
Allow logging stop with
STOP button
If this option is enabled, the LogTag can be stopped by pressing
the STOP button. The recorder will no longer take any readings,
however you can still review the day statistics and download the
logged results to a PC with LogTag Analyzer.
Allow reset of logger
with START button
If this option is enabled, TRID30-7F and TRID30-7R loggers can be
re-activated and used again for a new trip by pressing the START
button.
When re-activating the recorder, all previously recorded
readings and statistics will be erased!
When this option is enabled, a warning box will be displayed to
remind you this is enabled and can result in loss of data.
Show total summary
days collected
If this option is enabled, the LogTag shows the number of days for
which statistics have already been collected on the display.
Please refer to the
LogTag TRID30-7F product specification & guide
(http://www.logtagrecorders.com/products/pdf/LOGTAG_TRID30-7_PRODUCT_SPECIFIC
ATION.pdf) for a detailed description on the functionality of these products.
Содержание Analyzer
Страница 1: ...2 0...
Страница 23: ...Chapter 3 Quick Start Guide 23 Only flashes if no alert condition is present...
Страница 170: ......