OPERATION
6
59
IFC 050
www.krohne.com
09/2016 - 4002184003 - MA IFC 050 R03 en
Function
Setting / Description
C4 I/O HART
The dynamic variables are fixed to the following values:
PV: flow (volume flow or mass flow, depending on the selection of "measurement" in C1.1.4)
SV: counter 1
TV: counter 2
4V: operating time
C4.1 HART
HART
®
communication is activated or deactivated.
Select: on (HART
®
is activated; default setting) / off (HART
®
is disabled)
C4.2 address
Setting the address for multi-drop operation.
If "address 0" is selected, the current output is in normal operation. For all other
addresses, the current at the output is set to the 0% value.
C4.3 message
Entry of individual text.
C4.4 description
Entry of individual text.
C4.5 HART units
Unit change of dynamic variables in the display.
Select: break (return with
^
key) / HART
®
display (copies the settings for the
display units to the settings for the DVs) / load defaults (resets the DVs to the factory
defaults)
Function
Setting / Description
C6 device
C6.1 device info
C6.1.1 tag
Settable characters (max. 8 digits):
A
…
Z; a
…
z; 0
…
9; / - , .
C6.1.2 C number
CG number, cannot be changed; describes the signal converter version.
C6.1.3 device serial no.
Serial number of the system; cannot be changed.
C6.1.4 electronic serial no.
Serial number of the electronic assembly; cannot be changed.
C6.1.5 SW.REV.MS
Serial number of the circuit board, version number of the main software, production
date of the circuit board.
C6.1.6 Electronic Revision ER
Reference identification number, electronic revision and production date of the
device; includes all hardware and software changes.
C6.2 display
C6.2.1 language
Language selection depends on the device version.
C6.2.2 contrast
Adjust display contrast for extreme temperatures.
Setting: -9
…
0
…
+9
This change takes place immediately, not just when setting mode is exited!
C6.2.3 default display
Specification of the default display page that is returned to after a short delay
period.
Select: none (current page is always active) / 1. meas. page (shows this page) /
2. meas. page (shows this page) / status page (shows only status messages) /
graphic page (trend of the 1st measurement)
C6.2.4 magnet keys
For activating or deactivating the magnet keys.
Select: on (magnet keys are enabled) / off (magnet keys are disabled)