8
TECHNICAL DATA
82
IFC 050
www.krohne.com
09/2016 - 4002184003 - MA IFC 050 R03 en
8.5 Measuring accuracy
Every electromagnetic flowmeter is calibrated by direct volume comparison. The wet calibration
validates the performance of the flowmeter under reference conditions against accuracy limits.
The accuracy limits of electromagnetic flowmeters are typically the result of the combined effect
of linearity, zero point stability and calibration uncertainty.
Reference conditions
•
Medium: water
•
Temperature: +5...+35
°
C / +41...+95
°
F
•
Operating pressure: 0.1...5 barg / 1.5...72.5 psig
•
Inlet section:
≥
5 DN
•
Outlet section:
≥
2 DN
X [m/s]: flow velocity
Y [%]: Accuracy of measured value (mv)
DN [mm]
DN [inch]
Standard accuracy
1
Optimised accuracy
2
OPTIFLUX 1050
10
…
150
3/8
…
6
±
0.5% of mv
±
1 mm/s
±
0.25% of mv
±
1.5 mm/s
Extended calibration at
2 points
OPTIFLUX 2050
10
…
1200
3/8
…
48
OPTIFLUX 4050
OPTIFLUX 6050
10
…
150
3/8
…
6
WATERFLUX 3050
25...600
1...24
-