6
OPERATION
66
IFC 050
www.krohne.com
09/2016 - 4002184003 - MA IFC 050 R03 en
Information
Messages on the display Description
Actions
Status: I _ _ _ _ _
Information (current measurement OK)
I counter 1 stopped
This is counter 1. The counter has stopped.
If counter to continue counting, activate
"yes" in "C2.y.9 start counter".
I counter 2 stopped
This is counter 2. The counter has stopped.
I power fail
The device was not in operation for an
unknown period of time, because the power
was switched off. This message is for
information only.
Temporary power failure. Counters did not
run during it.
I over range display 1
1st line on page 1 (2) of display limited by
filter setting.
Menu display C4.3 and/or C4.4, select 1st or
2nd meas. page and increase values in
"C4.z.3 range" and/or "C4.z.4 limitation".
I over range display 2
I over range conductivity
The limits for the conductivity measurement
have been exceeded (>10000
µ
S/cm) or are
lower than permitted (<0.1
µ
S/cm).
If the flow sensor has been properly
connected and filled with medium, this does
not affect the flow measurement. The
conductivity measured values cannot be
used.
I backplane sensor
The data on the backplane are not usable
because they have been generated with an
incompatible version.
-
I backplane settings
The global settings on the backplane are not
usable because they have been generated
with an incompatible version.
-
I backplane difference
The data on the backplane differ from the
data in the display. If the data are usable, a
dialogue is indicated in the display.
-
I conductivity off
Conductivity measurement switched off.
Changing of settings in C1.3.1.
I empty pipe
1 or 2 measuring electrodes are not in
contact with the medium; measured value is
set to zero. No measurement possible.
Measuring pipe not filled, function
dependent on C1.3.2. Check installation. Or
electrodes completely insulated e.g. by oil
film. Clean!
I diagnosis channel off
Diagnosis value switched off.
Changing of settings in C1.3.17.