8
TECHNICAL DATA
76
IFC 050
www.krohne.com
09/2016 - 4002184003 - MA IFC 050 R03 en
Low flow cut off
Low flow cut off
Low flow cut off
Low flow cut off
Function
Switching point and hysteresis separately adjustable for each output, counter and
the display
Switching point
Set in increments of 0.1%.
0
…
20% (current output, frequency output) or 0...
±
9.999 m/s (pulse output)
Hysteresis
Set in increments of 0.1%.
0...5% (current output, frequency output) or 0...5 m/s (pulse output)
Time constant
Time constant
Time constant
Time constant
Function
The time constant corresponds to the elapsed time until 67% of the end value has
been reached according to a step function.
Settings
Set in increments of 0.1 seconds.
0
…
100 seconds
Status output / limit switch
Status output / limit switch
Status output / limit switch
Status output / limit switch
Function and settings
Adjustable as automatic measuring range conversion, display of flow direction,
counter overflow, error, switching point or empty pipe detection
Valve control with activated dosing function
Status and/or control: ON or OFF
Operating data
Operating data
Operating data
Operating data
Basic I/Os + Modbus
Basic I/Os + Modbus
Basic I/Os + Modbus
Basic I/Os + Modbus
Passive
Independent of connection polarity.
U
ext
≤
32 VDC
I
≤
100 mA
open:
I
≤
0.05 mA at U
ext
= 32 VDC
closed:
U
0
= 0.2 V at I
≤
10 mA
U
0
= 2 V at I
≤
100 mA
Modbus
Modbus
Modbus
Modbus
Description
Modbus RTU, Master / Slave, RS485
Address range
1
…
247
Broadcast
Supported with function code 16
Supported Baud rate
1200, 2400, 3600, 4800, 9600, 19200, 38400, 57600, 115200 Baud
Approvals and certificates
CE
This device fulfils the statutory requirements of the relevant EU directives.
The manufacturer certifies successful testing of the product by applying the CE
mark.
For full information of the EU directives & standards and the approved
certifications, please refer to the EU declaration or the manufacturer website.
Other standards and approvals
Other standards and approvals
Other standards and approvals
Other standards and approvals
Shock and vibration resistance
IEC 60068-2-3; EN 60068-2-6 and EN 60068-2-27; IEC 61298-3
NAMUR
NE 21, NE 43, NE 53