57
Transport and storage
BlueShadow Pump 80P Instructions V7685
algae formation, do not use pure water. Close all inlets and outlets with
blind screwings.
The pump has been flushed.
The pump has been switched off and disconnected from the power
supply.
Open-end wrench, size 10
Procedure
Process
1.
Remove the solvent tubing and seal all open connectors with cap
fit tings.
2.
Remove the pump from the HPLC system and seal the open
connec tor of the pressure sensor (isocratic version) or the mixer
(binary, ternary or quaternary version) with cap fittings.
The device can be stored under the ambient conditions described in
chapter (see chapter 11 on page 59).
9.4.2 Storing the pump head
If you want to store pump heads separately, proceed as follows:
The pump head has been dismounted (see chapter 8.5.1 on page 52).
Syringe
Flushing solution
Procedure
Process
1.
Fill the flushing solution into a syringe and inject into the capillary at
the pump head inlet. Leave it for 5 minutes.
2.
Flush with suitable purging solution.
3.
Fill the pump head with isopropanol.
4.
Seal the inlets and outles with cap fittings.
The pump head can be stored under the ambient conditions described in
chapter „Technical data“ on page 59.
Prerequisites
Tool
Storage conditions
Device defect
Residuals chemicals can damage the pump head if they are not
being removed before storage.
For reuse, fl ush the pump head and fi ll it with isopropanol.
Before storage, seal the inlets and outlets with cap fi ttings.
Prerequisite
Materials
Storage conditions