21
Installation
BlueShadow Pump 80P Instructions V7685
Connection
Function
6OC
TTL Output
Levels:
passive 5 V
active 0 V
Pulse:
5 V for at least 1000 ms
GND
Reference point of the voltage at the signal inputs.
Relay Contact
The contact is on a floating basis. Its setting depends
on the settings in the Control Unit or software.
Steady-rate signal:
passive = open relay contact
active = closed relay contact
Pulse:
Closed relay contact for at least 1000 ms
Permissible load of the relay contact: 1 A/ 24 V DC
Relay Contact
The contact is on a floating basis. Its setting depends
on the settings in the Control Unit or software.
Steady-rate signal:
passive = open relay contact
active = closed relay contact
Pulse:
Closed relay contact for at least 1000 ms
Permissible load of the relay contact: 1 A/ 24 V DC
Analog
GND
Reference point of the voltage at the signal inputs.
Analog out
Voltage range 0 – 5 V, scalable
Analog in
Power range 0 – 10 V
10 V according to maximum flow rate
Start GND
Reference point of the voltage at the signal inputs.
Start OUT
TTL Output
Levels:
passive 5 V
active 0 V
Output is active for 500 ms when a program starts.