49
Troubleshooting
BlueShadow Pump 80P Instructions V7685
System message
Solution
Program is running
Quit program or wait until program has been
completed.
S
Sum of compo-
nents is not 100
Change the entry.
T
The gradient com-
ponent is used!
First unload the program. The
setup
data can
only be edited when no program has been load-
ed or started.
this link is used in
WAKEUP
First quit or delete wakeup program (
WakeUp
)
and then edit or delete the link.
This program is
used in a link
First pause or delete the link, then edit or delete
data by means of the chromatography software.
this program is
used in WAKEUP
First quit or delete wakeup program (
WakeUp
)
and then edit or delete the data by means of the
chromatography software.
Time already exists
Correct the time entry.
Time table line is
empty
Edit the program line.
Too many lines in
program
Check the number of program lines. A maxi-
mum of 100 program lines are possible.
U
Unknown pump
head type!
Check the pump head.
Check whether a pump head with RFID rec-
ognition has been installed.
W
Wake up time
already passed!
Correct the entry for date or otherwise time.
Wrong Line
number
Change the line index in the
link
.
Wrong link
Check
Link
-Nr. Values from 1-10 are possible
Correct the entry.