13
Installation
BlueShadow Pump 80P Instructions V7685
Device defect
Changes of the environmental temperature cause condensation inside
the device.
Allow device to acclimate for 3 h before connecting to power supply
and taking into operation.
4.1.2 Power supply
For power supply, use the supplied power cable and power adapter to
meet the specifications which are described in the chapter Technical
Data. But inspect the provided power cable beforehand to ensure that it
is approved for your country. Replace defective power cables only with
accessories from KNAUER. Detachable power supply cables are not
allowed to be replaced with other cable types.
The electrical power supply at the installation site must be connected
directly to the nearest main power line.
The power must be free from ripple, residual current, voltage peaks
and electromagnetic interference.
The connectors for the mains voltage are grounded accordingly.
The device receives sufficient power with reserve capacity
The device is intended for use with AC power networks of 100-240 V.
Make sure that the power plug on the power supply (wall socket or
power strip) is always accessible, so that the device can be disconnect-
ed from the power supply.
4.1.3 Unpacking the device
Check the carton for damage caused during transportation.
Utility knife
Electronic defect
Electronic hazard when using an identically constructed power adapter
from another manufacturer.
Only use spare parts and accessories from KNAUER or a company
authorized by KNAUER.
Conditions
Power plug
Prerequisites
Tools
Bruising danger
Damage to the device by carrying or lifting it on protruding housing
parts. The device may fall and thus cause injuries.
Lift the device only centrally on the side of the housing.