26
Installation
BlueShadow Pump 80P Instructions V7685
1.
Use the patch cable to connect the router and the computer. Repeat
this step to connect all devices.
2.
Use the power supply to connect the router to the mains power
system.
4.5.7 Configuring the router
The router is preset at the factory. Information about address, user name
and password is noted in the router manual:
1.
To open the router configuration, start your Internet browser and
enter the IP address (not for all routers).
2.
Enter user name and password.
3.
Configure the router as DHCP server.
4.
In the router configuration, check the IP address range and make
changes if necessary.
Note:
If the IP address range should have been changed, make surte to
note his information on the router.
Once the router has assigned IP addresses to all devices, the chromatog-
raphy software can be used to remotely control the system.
4.5.8 Integrating the LAN into the company network
A network administrator can integrate the LAN into your company net-
work. In this case you use the WAN port of the router.
There is a patch cable for the connection.
1.
Check that the IP address range of the router and of the company
network do not overlap.
2.
In case of an overlap, change the IP address range of the router.
3.
Use the patch cable to connect the router WAN port to the company
network.
4.
Restart all devices, including the computer.
4.5.9 Controlling several systems separately in the LAN
Devices connected to a LAN communicate through ports, which are part
of the IP address. If more than one chromatography systems are connect-
ed to the same LAN and you plan on controlling them separately, you can
use different ports to avoid interference. Therefore, the port number for
each device must be changed and this same number must be entered
into the device configuration of the chromatography software. We recom-
mend to use the same port number for all devices in the same system.
Note:
The port is set to 10001 at the factory. You must use the same num-
bers in the device configuration of the chromatography software as in the
device, otherwise the connection fails.
1.
Find out port number and change it on the device.
2.
Enter the port number in the chromatography software.
The connection is established.
Procedure
Procedure
Result
Prerequisites
Procedure
Procedure
Result