Technical data
BlueShadow Pump 80P Instructions V7685
11. Technical data
11.1 Solvent delivery
Pump type
Dual-piston pump with two operating
pistons
Flow rate range
100 ml pump head:
—
0.01-100 ml/min
—
1-80 ml/min (recommended)
—
1-40 ml/min (continous working)
250 ml pump head:
—
0.01–250 ml/min
—
2.5-200 ml/min (recommended)
—
2.5-100 ml/min (continous
working)
500 ml pump head:
—
0.01–500 ml/min
—
5-400 ml/min (recommended)
—
5-200 ml/min (continous working)
1000 ml pump head:
—
1–1000 ml/min
—
10-800 ml/min (recommended)
—
10-400 ml/min (continous working)
Maximum pressure
100 ml pump head:
—
400 bar to 100 ml/min
250 ml pump head:
—
225 bar to 100 ml/min
—
Linear reduction: 225-200 bar of
100-150 ml/min
—
200 bar of 150-250 ml/min
500 ml pump head:
—
100 bar to 500 ml/min
1000 ml pump head:
—
75 bar to 350 ml/min
—
Linear reduction: 75-50 bar of
350-600 ml/min
—
50 bar of 600-1000 ml/min
Flow rate accuracy
+/- 2 % at 5 – 50 % of the flow range with
ethanol/water 10/90
Flow rate precision
< 0,1 % RSD, measured at 5 – 50 % of the
flow range with ethanol/water 10/90