12
Installation
BlueShadow Pump 80P Instructions V7685
4. Installation
This chapter describes all preparatory steps prior to start-up.
4.1 Preparations
4.1.1 Operating location
The intended use be ensured only if the requirements for ambient con-
ditions of the operating environment are met. You will find the ambient
conditions under Technical Data.
Device defect
The device overheats at exposure to sunlight and insufficient air circula-
tion. Device failures are very likely.
Set up the device in such a way that it is protected against exposure
to direct sunlight.
Leave room for air circulation: See paragraph „space requirements“.
At least 5 cm, if there is another device on one side.
At least 10 cm, if there are devices set up on both sides.
At least 15 cm to the cooler fan on the rear.
Make sure that the power plug on the power supply (wall socket or
power strip) is always accessible, so that the device can be disconnect-
ed from the power supply.
Position the device on a level surface.
Protect the device against direct exposure to sunlight.
Set up the device at a location not exposed to air drafts such as air
conditioning systems.
Do not set up the device near to other machines that cause floor
vibrations.
Keep the the devices away from high frequency sources. High frequen-
cies may compromise measuring values.
Note:
Make sure that the power plug on the rear of the device is always
accessible, so that the device can be disconnected from the power
supply.
Note:
Avoid electrostatic discharges at the pump head.
Observe the following when moving the device from a cooling room to a
normal laboratory:
Switch on the device before moving and leave switched on for ½ hour
to warm up.
As an alternative, allow the device to warm up in the laboratory for at
least 3 hours and do not connect to the power supply during this dura-
tion to avoid damages by condensed water.
Space requirements
General purpose
Change of locaction