48
Troubleshooting
BlueShadow Pump 80P Instructions V7685
System message
Solution
Invalid time table
index
Change the entry in the program line.
I2C failed for panel
Switch the device off and on. Inform the techni-
cal support in case the system message repeats
itself.
L
Link is loaded
First unload the
link
then change the
link
or
delete it.
Link is running
Wait until the
link
has been completed, then
change the
link
or delete it.
M
Maximum pressure!
System stopped
Reduce the pressure or adjust the upper
pressure limit.
Restart the system.
Memory error
Switch the device off and on. Inform the techni-
cal support in case the system message repeats
itself.
Minimum pressure!
System stopped
Increase the pressure or adjust the lower
pressure limit.
Restart the system.
Motor failure
Switch the device off and on. Inform the techni-
cal support in case the system message repeats
itself.
N
No link available
Create a
link
and edit it.
No link available,
pls edit link first
Create a
link
and edit it.
No time table to
start
Edit the data by means of the chromatography
software.
Non-existing com-
ponent is set to
non-0 value
Switch on the channel or edit the data using the
chromatography software.
Not enough space
to store link
Check the pump.
Check the number of program lines. A maxi-
mum of 100 program lines are possible.
Not enough space
to store program
Check the pump.
Check the number of program lines. A maxi-
mum of 100 program lines are possible.
Not supported
Change the entry.
P
Prg not compatible
with pump head!
Modify the program or replace the pump head.
Program does not
exist
Create and edit a program.