52
Maintenance and care
BlueShadow Pump 80P Instructions V7685
Flushing solution
Silicone tubing
Procedure
Process
Figure
1.
Immerse the solvent tubing
into the flushing solution.
2.
Plug a silicone tube onto the
venting nozzle
2
of the pres-
sure sensor.
3.
Open the venting screw
1
.
4.
Start the pump at a medium
flow rate.
Fig. 21:
Pressure sensor
Bring the pump into operation again.
8.5 Pump head
Depending on the requirements of the user, different pump heads are
used.
8.5.1 Dismounting the pump head
The pump head has been flushed.
Allen screwdriver
Note:
Loosen the opposite pairs of fastening screws of the pump head
evenly and alternately to prevent the pump pistons on the inside from
jamming.
Materials
Next step
Prerequisites
Tools
Life-threatening injuries
Health danger if getting in contact with toxic, infectious or radio-active
substances.
Before disposing of the device or sending it away for repair, you are
required to decontaminate the device in a technically correct manner.
1
2