5-4 Basic Operation
PLZ-4W
■
Displaying the elapsed time after the load is turned on
By default, the elapsed time after the load is turned on is not displayed.
If you select “1. Setup” | “1. Function” from the menu and set the Count Time set-
ting to on, the time from load on to load off can be displayed.
This function is useful when used in conjunction with the UVP in the discharge test
of batteries and capacitors.
•
For a description of changing the menu setup, see section 5.13, “Menu Setup.”
■
Turning off the load after a speci
fi
ed time elapses
By default, the load on timer is off.
If you select “1. Setup” | “4. Cut Off” from the menu and set the time in “Time,” the
load automatically turns off after a speci
fi
ed time elapses after the load is turned on.
When the load turns off, a pop-up window appears indicating the input voltage at
the time the load was turned off. This function is useful in discharge tests of batter-
ies and capacitors.
•
For a description of changing the menu setup, see section 5.13, “Menu Setup.”
■
Turning on or off the load externally
You can also turn on or off the load using an external signal through a relay or other
similar methods.
•
For a description of external control, see section 6.8.7, “External Control of
Load On and Load Off.”
■
Starting up the input current of the PLZ-4W responds to follow
the rise time of the output voltage
In the constant current (CC) mode and the constant resistance (CR) mode, you can
set to start up slowly of the input current of the load (soft start).
•
For a description of soft-start, see section 5.10, “Soft start.”