6-28 Applied Operation
PLZ-4W
A program is a collection of steps (execution unit). When program 11 is executed
(RUN), steps are executed one by one in ascending order from step number 0001.
The completion of the last step signi
fi
es the end of one program execution.
• Step
Current
Current for the step
STEP
Step number
T
Specify the step when you wish to output a trigger signal using T.
• TRIG (trigger output)
TRIG sets the presence or absence of
trigger output. When turned on, a trigger
signal is output from the TRIG OUT
connector on the front panel at the same
time as the step is executed.
• FILL function
This function automatically sets the current values of steps in between two sepa-
rated steps for which you specify the current values.
Fig. 6-18
Setup example using the FILL function
• DATA1
Current value at the start step (0.000 to 100 % of the range setting)
• DATA2
Current value at the end step (0.000 to 100 % of the range setting)
• START
Start step number
• STOP
End step number
In Fig. 6-18, the same waveform is repeated twice. DATA1 and DATA2 of the two
waveforms are set to the same values, respectively, for each two-point interval. The
values between two points are automatically assigned.
Step n-1
Step n
TRIG OUT
(Example) Set trigger output at step
number n (TRIG ON)
Fig. 6-17
TRIG sequence example
10.00
A
STEP:0014
M1
DATA1: 0.164
A
DATA2: 1.164
A
START: 0016
STOP :
ENTER
FILL
0100
10.00
A
STEP:0014
M1
DATA1: 0.164
A
DATA2: 1.164
A
START: 0101
STOP :
ENTER
FILL
0200
Current
value
1.164 A
0.164 A
Step
16
100 101
200
Steps 16 to 100
Steps 101 to 200
(DATA2)
(DATA1)
(START) (STOP) (START) (STOP)