PLZ-4W
Maintenance and Calibration 8-7
8.3.3
Calibration Procedure
Entering the calibration screen and selecting Calibration
1.
Press the MENU (SHIFT+SET/VSET) key.
The menu screen is displayed.
2.
Use the CURSOR key to select 3. Calibration.
3. Calibration is highlighted.
3.
Press the ENTER key.
The Calibration screen appears.
Fig. 8-6
Calibration screen
■
Alarm
If an alarm occurs while the calibration is in progress, the load turns off along with
a beeping alarm. Press the ENTER key after clearing the problem which was caused
to activate an alarm, then the Calibration screen appears on the display and a beep-
ing alarm stops. In this case, go back to the step from the beginning of the alarm
occurd calibration number.
1.
CC
(
Low
)
2. CC
(
Mid
)
4. CV 15V
5. CV 150V
PREV
NEXT
3. CC
(
Hi
g
h
)
Calibration
Calibration
number