8-4 Maintenance and Calibration
PLZ-4W
8.2
Con
fi
rming status of the fuse
To increase security of the product, the fuse are installed on each element of the
load device. Those fuses protects from the effect to the connected device at mini-
mum level causing by the failure or incorrect use of the product. If the setting cur-
rent is not
fl
owing properly, the fuse may have blown off. Please check the
following procedure.
1.
Apply the current
fl
ow while the loads are connected.
2.
Remove the whole louvers on the front panel, and check if the LED
lights in red.
If the LED lights, the fuse may have blown off. In this case, contact your
Kikusui distributor/agent for request of service.
Check for the input voltage and the input current within the rated value. If the mea-
sured value is out of rated range, the LED may lights on even the fuse does not blow
off.