3-10 For First Time Users
PLZ-4W
2.
Select CC mode.
Press the MODE key to show the operation mode pop-up menu.
If CC is not highlighted, press the MODE key repetitively until CC is high-
lighted.
When the pop-up menu closes, check that CC appears on the display.
3.
Select the current range.
Press the RANGE key to show the current range pop-up menu.
Each time you press the RANGE key while the menu is displayed, the current
range switches in the order L (LOW), M (MIDDLE), and H (HIGH). When the
desired range is highlighted, stop pressing the key.
Along with L, M, or H, the full scale value of the respective range is displayed.
This value varies depending on the model.
4.
Select the voltage range.
Press the VRANGE (SHIFT+RANGE) key to show the voltage range pop-up
menu.
Each time you press the VRANGE key while the menu is displayed, the volt-
age range switches between 15 V and 150 V. When the desired range is high-
lighted, stop pressing the key.
5.
Set the current value.
Press the SET/VSET key to illuminate the SET/VSET key.
While viewing the display, turn the rotary knob.
You can press the rotary knob to switch between coarse adjustment and
fi
ne
adjustment. First, use coarse adjustment to roughly set the value, and then
switch to
fi
ne adjustment to adjust the value precisely.
6.
Set the OPP value.
Press the OPP/OCP key to illuminate the OPP/OCP key and show the OPP
value on the display.
While viewing the display, turn the rotary knob.
You can press the rotary knob to switch between coarse adjustment and
fi
ne
adjustment. First, use coarse adjustment to roughly set the value, and then
switch to
fi
ne adjustment to adjust the value precisely.
7.
Turn on the load.
Press the LOAD key to illuminate the LOAD LED and allow the current to
fl
ow.
At this point, the display shows the measured values (current, voltage, and
power) at the load input terminal.
8.
Change the current setting.
Turn the rotary knob while the load is turned on to change the current setting.
However, the current value cannot be set greater than the maximum value of
the selected range.
9.
Turn off the load.
Press the LOAD key to turn off the LOAD LED.
0.00
mA
0.000
V
0.000
W
CC
SLEWRATE
0.000
A
0.01
mA/
μ
s
SET
H 33A
150V
CP
CC
CR
CV
0.00
mA
0.000
V
0.000
W
CC
SLEWRATE
1.000
A
0.01
mA/
μ
s
SET
H 33A
150V
L
M
0.33A
3.3A
33A
H
0.00
mA
0.000
V
0.000
W
CC
0.000
A
SET
H 33A
150V
181.5
W