6-32 Applied Operation
PLZ-4W
• No.
Program number (11)
• Memo
Memo (PLZ164W)
• Range
Current range-voltage range (33A-15V)
• Loop
Number of program loops (3)
• Last Load
Load condition at the end (ON)
• Last Set
Current value at the end (1 A)
• RPTSTEP
Last step per loop (15)
• TIME BASE
Step execution time (1 ms)
• Current setting
Steps 1 to 5: 10 A
Step 5: 5 A
Steps 7 to 10: 10 A
Step 11: 5 A
Steps 12 to 15: 10 A
• Trigger output
Step 6
Creating the program
1.
Select sequence edit.
Press the EDIT (SHIFT+A) key to display the program edit screen.
■
Setting the operation of program 11
2.
Select the program number.
Turn the rotary knob to set the program number.
Select 11, because you can only select 11 for a fast sequence. After setting the
program number, press the CURSOR key to move to the next item.
3.
Enter a memo.
When adding a memo, turn the rotary knob to enter the characters.
In program 11, enter “Program11”. After entering the memo, move the
CURSOR key to move to the next item.
4.
Select the operation mode.
Turn the rotary knob to show the operation mode to be executed.
Select constant current mode “FCC.” After setting the operation mode, move
the CURSOR key to move to the next item.
5.
Sets the range.
Turn the rotary knob to select the current range and voltage range.
Set 33 A - 15 V. After setting the range, move the CURSOR key to move to
the next item.
SEQ
PREV
NEXT
No. :
11
Memo :
Mode : FCC
Loop : 0003
Ran
g
e : 33A , 15V