PLZ-4W
Applied Operation 6-33
6.
Set the number of program loops.
Turn the rotary knob to show the number of program loops.
Set 0003 to specify 3 loops.
At the last line, move the CURSOR key to move to the next item.
7.
Set the load condition after the end of the program.
Turn the rotary knob to select the load condition.
Since the load is turned on at the end of program 11, select ON.
After setting the load condition, move the CURSOR key to move to the next
item.
8.
Set the current setting after the end of the program.
Turn the rotary knob to set the current value.
Since the load is turned on at the end of program 11, select 1 A.
After selecting the current setting, move the CURSOR key to move to the
next item.
9.
Set the last step per loop for program 11.
Turn the rotary knob to set the step number.
Set 15.
After setting the step number, move the CURSOR key to move to the next
item.
10. Set the step execution time.
Turn the rotary knob to set the step execution time.
Set 1 ms.
After setting the step execution time, press the ENTER key to move to the next
edit screen.
■
Setting the current value and trigger output of each step
We will set the current value and trigger output of each step of program 11.
Here, we will set the current value
fi
rst followed by the trigger output.
Fig. 6-20
Step setup screen
SEQ
PREV
NEXT
No. :
11
Memo :
Mode : FCC
Loop : 0003
Ran
g
e : 33A , 15V
SEQ
PREV
NEXT
Last Load : ON
Last Set : 1.000
A
RPTSTEP : 15
TIME BASE : 1ms
10.00
A
STEP:0014
M1
Current value
Step number
Trigger output
Magnification
Waveform display
Specify the item using
the and keys.