5-28 Basic Operation
PLZ-4W
5.14 Initialization
The backup function of the PLZ-4W retains the current settings, menu settings, and
memory contents (ABC preset memories and setup memory) even when the
POWER switch is turned off. You can initialize the PLZ-4W settings to factory
default by carrying out the following operation.
•
For a description of the ABC preset memories, see section 6.1, “ABC preset
memories.” For a description of the setup memory, see section 6.2, “Setup
Memory.”
•
For a description of the default menu settings, see Table 5-3.
• Use caution because the contents of ABC preset memories and the setup
memory are also cleared when the PLZ-4W is initialized.
■
Initializing the PLZ-4W
1.
While holding down the ENTER key, turn on the POWER switch.
SET CLR appears when the display turns on.
Release the ENTER key when the next information appears.
The PLZ-4W settings are initialized.
Table 5-3
Factory default settings
Item
Panel settings
Setup memory settings
(all 100 sets)
OCP value
Maximum value
Minimum value
OPP value
Maximum value
Minimum value
UVP value
OFF
OFF
Current
0 A
Full scale of H range
Conductance
0 S
Full scale of H range
Voltage
Minimum value
Full scale of H range
Wattage
0 W
Full scale of H range
Current range
H
H
Voltage range
150 V
150 V
Load on/off
Load off
Load off
Operation mode
CC
CC
Slew rate
Minimum value of H range
Maximum value of H range
ABC preset memories
Settings above in each mode
Settings above in each mode
Menu setup
CAUTION