PLZ-4W
Applied Operation 6-39
6.8
External Control
6.8.1
Overview and Precaution of External Control
The settings in each operation mode normally use the internal reference signal. In
external control, this reference signal is supplied externally. The external signal is
either voltage (voltage control) or resistance (resistance control).
External control can control the settings from 0 % to 100 % of the selected range in
the CC, CR, CP, or CV mode.
Table 6-6
External control using voltage or resistance
There are other external controls as indicated below.
Table 6-7
Other external controls
Control method
Operation mode
Description
Voltage
CC, CP, CV
Change of 0 V to 10 V produces a change of 0 % to
100 % of the rated value of the range
CR
Change of 0 V to 10 V produces a change of maxi-
mum value to minimum value of the range
Resistance
(proportional)
CC, CP, CV
Change of 0
Ω
to 10 k
Ω
produces a change of 0
% to 100 % of the rated value of the range
CR
Change of
0
Ω
to
10 k
Ω
produces a change of
maximum value to minimum value of the range
Resistance
(inverse proportional)
CC, CP, CV
Change of 10 k
Ω
to 0
Ω
produces a change of 0 % to
100 % of the rated value of the range
CR
Change of 10 k
Ω
to 0
Ω
produces a change of maxi-
mum value to minimum value of the range
Function
Description
Load on/off
Turns on/off the load and monitor the load
Current range switching
Controls the range of each operation mode or monitor the range
Trigger input
Clear the pause in sequence operation
Alarm input
Forcibly activates an alarm
Current monitor
Monitors the input current
Short signal
Relay contact output
Содержание PLZ-4W Series
Страница 10: ...VIII Contents PLZ 4W ...
Страница 20: ...1 10 General Information PLZ 4W ...
Страница 38: ...2 18 Installation and Preparation PLZ 4W ...
Страница 90: ...5 30 Basic Operation PLZ 4W ...
Страница 192: ...9 10 Specifications PLZ 4W ...
Страница 214: ...A 22 Appendix PLZ 4W ...