4-2
8
KFM2151
Remote Control
4.6.1
IEEE488.2 Register Model
Status byte register
The status byte register stores STB and RQS (MSS) messages as defined by the
IEEE488.1 standard. The status byte register can be read using IEEE488.1 serial
polling or IEEE488.2 common command *STB?.
When serial polling is carried out, bit 6 responds with the request service (RSQ).
The status byte value is not changed by serial polling.
The *STB? query makes the device transmit the contents of the status byte register
and the master summary status (MSS) summary message.
The *STB? query does not change the status byte, MSS, and RQS.
Ta
b
le 4-4
Stat
u
s
b
yte s
u
mmary register
Bit
Bit
W
eight
Bit
N
ame
Description
0
1
Reser
v
ed
Reser
v
ed for f
u
t
u
re
u
se
b
y the IEEE. The
b
it
v
al
u
e is
notified as zero.
1
2
Reser
v
ed
2
4
Error/E
v
ent Q
u
e
u
e (EEQ)
If data exists in the error or e
v
ent q
u
e
u
e, this
b
it is
set to tr
u
e.
3
8
Q
u
estiona
b
le Stat
u
s Register
(QUES)
This
b
it is set to tr
u
e
w
hen a
b
it is set in the
QUEStiona
b
le e
v
ent stat
u
s register and the
corresponding
b
it in the QUEStiona
b
le stat
u
s ena
b
le
register is tr
u
e.
4
16
Message A
v
aila
b
le (MA
V
)
This
b
it is set to tr
u
e
w
hen a req
u
est is recei
v
ed from
the digital programming interface and the KFM2151
is ready to o
u
tp
u
t the data
b
yte.
5
32
Standard E
v
ent Stat
u
s Bit
S
u
mmary (ESB)
This
b
it is set to tr
u
e
w
hen a
b
it is set in the e
v
ent
stat
u
s register.
6
64
Req
u
est Ser
v
ice (RQS)
This
b
it is set to tr
u
e
w
hen a
b
it is set in the ser
v
ice
req
u
est ena
b
le register, and the corresponding
b
it
exists in the stat
u
s
b
yte.
Master Stat
u
s S
u
mmary
(MSS)
Set
b
y one of the follo
w
ing:
Stat
u
s
b
yte
b
it 0 and ser
v
ice req
u
est ena
b
le register
b
it 0.
Stat
u
s
b
yte
b
it 1 and ser
v
ice req
u
est ena
b
le register
b
it 1.
Stat
u
s
b
yte
b
it 2 and ser
v
ice req
u
est ena
b
le register
b
it 2.
Stat
u
s
b
yte
b
it 3 and ser
v
ice req
u
est ena
b
le register
b
it 3.
Stat
u
s
b
yte
b
it 4 and ser
v
ice req
u
est ena
b
le register
b
it 4.
Stat
u
s
b
yte
b
it 5 and ser
v
ice req
u
est ena
b
le register
b
it 5.
Stat
u
s
b
yte
b
it 7 and ser
v
ice req
u
est ena
b
le register
b
it 7.
7
12
8
Operation Stat
u
s Register
(OPER)
This
b
it is set to tr
u
e
w
hen a
b
it is set in the
OPERation e
v
ent stat
u
s register and the
corresponding
b
it in the OPERation stat
u
s ena
b
le
register is set.
8
-15
N
ot Used
–
Содержание KFM2151
Страница 1: ...OPERATION MANUAL FC Scanner Part No Z1 003 382 IB010911 May 2006 KFM2151...
Страница 4: ...ii KFM2151...
Страница 14: ...xii KFM2151 Contents...
Страница 19: ...General Description This chapter gives a general description and introduces the features of the KFM2151...
Страница 23: ...Installation and Preparation This chapter describes the procedures from unpacking to installation...
Страница 35: ...Operation This chapter mainly describes various setup procedures...
Страница 54: ...3 20 KFM2151 Operation...
Страница 81: ...KFM2151 4 27 Remote Control Fig 4 3 Status register SCPI Standard 1999 0 Volume1 fig 9 1...
Страница 100: ...4 46 KFM2151 Remote Control...
Страница 101: ...Maintenance This chapter covers maintenance of the KFM2151 including cleaning inspecting calibrating and troubleshooting...
Страница 105: ...Specifications This chapter contains the electrical and mechanical specifications of the KFM2151...
Страница 110: ...I 2 KFM2151 Index String 4 8 T terminator 4 7 U unit number setting of 3 13 V voltage protection 3 5...
Страница 111: ......
Страница 112: ...FC SCANNER KFM2151 OPERATION MANUAL...