KFM2151
2-9
Installation and Preparation
2.5
Connecting the SENSING terminal and the
fuel cell
Use the screwless terminal connector (8 poles) provided. One block contains four
channels of terminals. Connect the SENSING terminal directly to the separator in
order from cell 1.
Fig.2-
8
W
iring the SE
N
SI
N
G terminal
If you want to measure the performance at the electrode terminal, connect to it.
Use AWG24 wires. The recommended amount of covering to be stripped is
approximately 10 mm. Using the strip gauge shown in Fig.2-9 will ensure proper
work.
•
Check the rated temperature of the wires that you are using. Pay close attention
for fuel cells that operate at high temperatures. Even for fuel cells that run at
relatively low temperatures such as a PEFC, the collecting electrode where the
wire is connected may be hot.
Fig.2-9
Connection to the f
u
el cell
1.
Using a scre
w
dri
v
er, insert the
w
ire into the scre
w
less terminal
connector (
8
poles) pro
v
ided as sho
w
n in Fig.2-9.
Fuel cell
SENSING terminal
Cell 1
Cell 2
Cell 3
Wire
1
1
2
2
3
3
4
4
+
-
Fuel cell
Load cable
Twist as much as possible
Twist
Screwless terminal
connector (8 poles)
Wire insertion hole
Insert the wire by pressing with
the tip of a screwdriver.
Return it to the original position
after inserting the wire to fix
the wire in place
Insert into the input
(SENSING) terminal of
the KFM2151
Strip gauge
10 mm
Содержание KFM2151
Страница 1: ...OPERATION MANUAL FC Scanner Part No Z1 003 382 IB010911 May 2006 KFM2151...
Страница 4: ...ii KFM2151...
Страница 14: ...xii KFM2151 Contents...
Страница 19: ...General Description This chapter gives a general description and introduces the features of the KFM2151...
Страница 23: ...Installation and Preparation This chapter describes the procedures from unpacking to installation...
Страница 35: ...Operation This chapter mainly describes various setup procedures...
Страница 54: ...3 20 KFM2151 Operation...
Страница 81: ...KFM2151 4 27 Remote Control Fig 4 3 Status register SCPI Standard 1999 0 Volume1 fig 9 1...
Страница 100: ...4 46 KFM2151 Remote Control...
Страница 101: ...Maintenance This chapter covers maintenance of the KFM2151 including cleaning inspecting calibrating and troubleshooting...
Страница 105: ...Specifications This chapter contains the electrical and mechanical specifications of the KFM2151...
Страница 110: ...I 2 KFM2151 Index String 4 8 T terminator 4 7 U unit number setting of 3 13 V voltage protection 3 5...
Страница 111: ......
Страница 112: ...FC SCANNER KFM2151 OPERATION MANUAL...