4-22
KFM2151
Remote Control
4.5.5
Trigger Function
INIT
Initiates the voltage scanning. INIT discards the measured data that is already being
held immediately before performing the measurement. If an FETC? query is sent
immediately after INIT, measured data is returned after waiting for the measurement
to complete.
Command
INITiate[:IMMediate][:ALL]
ABOR
Aborts the scanning currently in progress. Does nothing if the measurement is not in
progress.
The condition of the KFM2151 after the power is turned on or immediately after
switching from local to remote mode is the same as the condition after ABOR is
sent. If ABOR is sent while measurement is in progress, the measured data remains
void. If ABOR is sent when the status has not be initialized and the measured data
being held is valid, the measured data is never discarded.
Command
ABORt
*RST
Aborts voltage scanning and discards the measured data being held. Therefore,
*RST;:FETC:VOLT? results in an SCPI error (-230, "Data corrupt or stale"),
because a query is being made when there is no valid measured data and acquisition
of new data cannot be expected.
Command
*RST
Содержание KFM2151
Страница 1: ...OPERATION MANUAL FC Scanner Part No Z1 003 382 IB010911 May 2006 KFM2151...
Страница 4: ...ii KFM2151...
Страница 14: ...xii KFM2151 Contents...
Страница 19: ...General Description This chapter gives a general description and introduces the features of the KFM2151...
Страница 23: ...Installation and Preparation This chapter describes the procedures from unpacking to installation...
Страница 35: ...Operation This chapter mainly describes various setup procedures...
Страница 54: ...3 20 KFM2151 Operation...
Страница 81: ...KFM2151 4 27 Remote Control Fig 4 3 Status register SCPI Standard 1999 0 Volume1 fig 9 1...
Страница 100: ...4 46 KFM2151 Remote Control...
Страница 101: ...Maintenance This chapter covers maintenance of the KFM2151 including cleaning inspecting calibrating and troubleshooting...
Страница 105: ...Specifications This chapter contains the electrical and mechanical specifications of the KFM2151...
Страница 110: ...I 2 KFM2151 Index String 4 8 T terminator 4 7 U unit number setting of 3 13 V voltage protection 3 5...
Страница 111: ......
Страница 112: ...FC SCANNER KFM2151 OPERATION MANUAL...