3.5
Logical Channel Configuration
Logical channels and SENSING terminal
The channels on the KFM2151 are logical channels. You can assign arbitrary
SENSING terminal numbers to be assigned to channels (logical channels). Up to 32
logical channels can be set separately for impedance and voltage.
On the SENSING terminal, TERM+ is high potential and TERM- is low potential.
Isolated from the chassis electric potential (withstand voltage of 250 V). The input
impedance is approximately 1 M
Ω.
Because the channels are logical channels, channel numbers can be changed without
having to change the connection to the measurement point. It is also possible to
assign the same SENSING terminal number to multiple logical channels.
By factory default, the SENSING terminal numbers and logical channel numbers
match. In the connection shown in Fig.3-5, channel 1 scans cell 1.
Fig.3-5
Logical channel config
u
ration (factory defa
u
lt condition)
Here, we change the configuration of channel 1. The SENSING terminals are
assigned to the voltage and impedance logical channels as shown below.
Assign TERM+3 and TERM-2 to voltage channel 1. The negative terminal of cell 2
and the positive terminal of cell 3 are assigned, and the series voltage of cells 2 and
3 is scanned.
Assign TERM+2 and TERM-1 to impedance channel 1. The negative terminal of
cell 1 and the positive terminal of cell 2 are assigned, and the series impedance of
cells 1 and 2 is scanned.
Fig.3-6
Logical channel config
u
ration (example in
w
hich channel
1 is changed)
Fuel cell
SENSING
terminal
Cell 1
Cell 2
Cell 3
Wire
Channel 1
By factory default, the SENSING terminal
numbers and voltage/impedance channel
numbers match.
Fuel cell
SENSING
terminal
Cell 1
Cell 2
Cell 3
Wire
Impedance channel 1
(TERM-1, TERM+2)
Voltage channel 1
(TERM-2, TERM+3)
Configuration of channel 1 whose
voltage and impedance have
been changed
Содержание KFM2151
Страница 1: ...OPERATION MANUAL FC Scanner Part No Z1 003 382 IB010911 May 2006 KFM2151...
Страница 4: ...ii KFM2151...
Страница 14: ...xii KFM2151 Contents...
Страница 19: ...General Description This chapter gives a general description and introduces the features of the KFM2151...
Страница 23: ...Installation and Preparation This chapter describes the procedures from unpacking to installation...
Страница 35: ...Operation This chapter mainly describes various setup procedures...
Страница 54: ...3 20 KFM2151 Operation...
Страница 81: ...KFM2151 4 27 Remote Control Fig 4 3 Status register SCPI Standard 1999 0 Volume1 fig 9 1...
Страница 100: ...4 46 KFM2151 Remote Control...
Страница 101: ...Maintenance This chapter covers maintenance of the KFM2151 including cleaning inspecting calibrating and troubleshooting...
Страница 105: ...Specifications This chapter contains the electrical and mechanical specifications of the KFM2151...
Страница 110: ...I 2 KFM2151 Index String 4 8 T terminator 4 7 U unit number setting of 3 13 V voltage protection 3 5...
Страница 111: ......
Страница 112: ...FC SCANNER KFM2151 OPERATION MANUAL...