Measurement Function
■
Difference between SCALar and ARRay
The second level node of FETC?, READ?, and MEAS? can be [SCALar] or ARRay.
[SCALar] is the default node that can be omitted and returns the measured data of
the channel specified by CHAN:VOLT:NSEL.
ARRay returns the measured data of all logical channels that are set as scanned
sources for voltage scanning in comma-separated format.
■
Difference between FETC, READ, and MEAS
FETC acquires data that has been measured already without starting the
measurement operation. If you use the FETC? query without initiating the voltage
scanning operation after the measured data has been discarded, an SCPI error (-230,
"Data corrupt or stale") occurs. This is because there is no valid measured data, and
the acquisition of measured data cannot be expected. Be sure to use this command
with the INIT command.
READ starts a new measurement operation and acquires the resultant data. The
READ command is a combination of INIT and FETC.
MEAS resets the settings to default, starts a new measurement operation, and
acquires the resultant data. The MEAS command is a combination of ABOR,
CONF, and READ.
FETC
FETC:ARR
Queries the voltage measurement data.
Command
FETCh[:SCALaar][:VOLTage[:DC]]?
FETCh:ARRay[:VOLTage[:DC]]?
Response
Returns the measured data in the <NR3> form in response to FETC?.
Returns the measured data in the <NR3>, <NR3>... form in response to
FETC:ARR?.
Unit:
V
Содержание KFM2151
Страница 1: ...OPERATION MANUAL FC Scanner Part No Z1 003 382 IB010911 May 2006 KFM2151...
Страница 4: ...ii KFM2151...
Страница 14: ...xii KFM2151 Contents...
Страница 19: ...General Description This chapter gives a general description and introduces the features of the KFM2151...
Страница 23: ...Installation and Preparation This chapter describes the procedures from unpacking to installation...
Страница 35: ...Operation This chapter mainly describes various setup procedures...
Страница 54: ...3 20 KFM2151 Operation...
Страница 81: ...KFM2151 4 27 Remote Control Fig 4 3 Status register SCPI Standard 1999 0 Volume1 fig 9 1...
Страница 100: ...4 46 KFM2151 Remote Control...
Страница 101: ...Maintenance This chapter covers maintenance of the KFM2151 including cleaning inspecting calibrating and troubleshooting...
Страница 105: ...Specifications This chapter contains the electrical and mechanical specifications of the KFM2151...
Страница 110: ...I 2 KFM2151 Index String 4 8 T terminator 4 7 U unit number setting of 3 13 V voltage protection 3 5...
Страница 111: ......
Страница 112: ...FC SCANNER KFM2151 OPERATION MANUAL...