Values for 34470A DMM
Input
Error from Nominal
Current Function
[1]
Range
Quick
Check
24 hours 90 days
1 year
2 years
2 A
DC Current
3 A
± 4.2 mA
± 4.6 mA
± 5.2 mA
1 A
1 A
Q
± 560 µA ± 800 µA ± 900 µA ± 1.1 mA
100 mA
100 mA
± 14 µA
± 35 µA
± 55 µA
± 65 µA
10 mA
10 mA
Q
± 2.7 µA
± 5 µA
± 7 µA
± 8 µA
1 mA
1 mA
± 0.1 µA ± 0.35 µA ± 0.55 µA ± 0.65 µA
[1]Select 10 PLC.
Q
: Quick performance verification test point.
Values for 34470A DMM
Input
Error from Nominal
Voltage Function
[1]
Range
Quick
Check
24 hours 90 days
1 year
2 years
100 mV
DC Volts
100 mV
± 6 µV
± 7.5 µV
± 8 µV
-100 mV
1 V
1 V
± 14 µV
± 19 µV
± 24 µV
± 29 µV
-1 V
4 V
10 V
± 52 µV
± 72 µV
± 84 µV ± 100 µV
10 V
Q
± 100 µV ± 150 µV ± 180 µV ± 220 µV
-10 V
100 V
100 V
± 2.6 mV ± 3.8 mV ± 4.4 mV ± 4.6 mV
-100 V
Q
1000 V
1000 V
± 36 mV
± 48 mV ± 54 mV ± 56 mV
-500 V
± 16 mV
± 22 mV ± 25 mV ± 36 mV
[1]Select 10 PLC.
Q
: Quick performance verification test point.
Values for 34470A DMM
Input
Error from Nominal
Resistance Function
[1]
Range
Quick
Check
24 hours
90 days
1 year
2 years
100 Ω
4-wire Ohms
100 Ω
± 6 mΩ
± 9 mΩ
± 10 mΩ
± 11 mΩ
1 kΩ
1 kΩ
Q
± 25 mΩ
± 35 mΩ
± 45 mΩ
± 55 mΩ
10 kΩ
10 kΩ
± 250 mΩ ± 350 mΩ ± 450 mΩ ± 550 mΩ
100 kΩ
100 kΩ
± 2.5 Ω
± 3.5 Ω
± 4.5 Ω
± 5.5 Ω
1 MΩ
1 MΩ
± 25 Ω
± 65 Ω
± 75 Ω
± 85 Ω
10 MΩ
10 MΩ
± 1.1 kΩ
± 2.1 kΩ
± 2.6 kΩ
± 3.1 kΩ
10 MΩ
2-wire Ohms
10 MΩ
Q
100 MΩ
100 MΩ
± 101 kΩ
± 201 kΩ
± 301 kΩ
± 401 kΩ
[1]Select 10 PLC.
Q
: Quick performance verification test point.
Performance Verification
514
Keysight True
volt
Series Operating and Service Guide
Содержание 34460A
Страница 1: ...Operating and Service Guide Keysight Truevolt Series Digital Multimeters ...
Страница 183: ...Measurement Tutorial 184 Keysight Truevolt Series Operating and Service Guide ...
Страница 385: ...STATus Subsystem SCPI Programming Reference 386 Keysight Truevolt Series Operating and Service Guide ...
Страница 541: ...Next Procedure Calibration Adjustment Procedures 542 Keysight Truevolt Series Operating and Service Guide ...
Страница 556: ...Next Procedure Keysight Truevolt Series Operating and Service Guide 557 Calibration Adjustment Procedures ...