Pin
Location Detail
Low reference: large metal shield
U1006 (large tab on right side as
shown in picture):
1.2 V ± 5%
Low reference: large metal shield
U1004 pin 3:
5 V ± 5%
Low reference: large metal shield
U301:
pin 2: +16.8 V ± 5%
pin 3: -16.4 V ± 5%
Keysight True
volt
Series Operating and Service Guide
481
Service and Repair
Содержание 34460A
Страница 1: ...Operating and Service Guide Keysight Truevolt Series Digital Multimeters ...
Страница 183: ...Measurement Tutorial 184 Keysight Truevolt Series Operating and Service Guide ...
Страница 385: ...STATus Subsystem SCPI Programming Reference 386 Keysight Truevolt Series Operating and Service Guide ...
Страница 541: ...Next Procedure Calibration Adjustment Procedures 542 Keysight Truevolt Series Operating and Service Guide ...
Страница 556: ...Next Procedure Keysight Truevolt Series Operating and Service Guide 557 Calibration Adjustment Procedures ...