![Keysight 34460A Скачать руководство пользователя страница 419](http://html1.mh-extra.com/html/keysight/34460a/34460a_operating-and-service-manual_1961230419.webp)
SYSTem:LOCK:NAME?
Returns the I/O interface in use by the querying computer.
Parameter
Typical Return
(none)
"LAN169.254.149.35"
See
l
After determining the interface that you are using, use
to determine which
interface, if any, has the lock.
l
Returns "USB", "VXI11", "GPIB", "LAN<
IP Address
>", or “WEB<IP Address>” indicating the I/O inter-
face being used by the querying computer.
SYSTem:LOCK:OWNer?
Returns the I/O interface that currently has a lock.
Parameter
Typical Return
(none)
"LAN169.254.149.35"
See
l
When a lock is active, Bit 10 in the Standard Operation Register will be set (
). When the lock is released on all I/O interfaces, this bit will be cleared.
l
Returns "USB", "VXI11", "GPIB", "LAN<
IP Address
>", or “WEB<IP Address>” indicating the I/O inter-
face that currently has a lock. If no interface has a lock, "NONE" is returned.
SYSTem:LOCK:RELease
Decrements the lock count by 1 and may release the I/O interface from which the command is executed.
Parameter
Typical Return
(none)
(none)
See
l
When a lock is active, Bit 10 in the Standard Operation Register will be set (
). When the lock is released on all I/O interfaces, this bit will be cleared.
SCPI Programming Reference
420
Keysight True
volt
Series Operating and Service Guide
Содержание 34460A
Страница 1: ...Operating and Service Guide Keysight Truevolt Series Digital Multimeters ...
Страница 183: ...Measurement Tutorial 184 Keysight Truevolt Series Operating and Service Guide ...
Страница 385: ...STATus Subsystem SCPI Programming Reference 386 Keysight Truevolt Series Operating and Service Guide ...
Страница 541: ...Next Procedure Calibration Adjustment Procedures 542 Keysight Truevolt Series Operating and Service Guide ...
Страница 556: ...Next Procedure Keysight Truevolt Series Operating and Service Guide 557 Calibration Adjustment Procedures ...