Utility - Autocalibration (34465A/70A only)
Autocalibration compensates for drift caused by time and temperature change in DC voltage and res-
istance measurements, and adjusts some aspects of other measurement functions.
To quickly perform autocalibration without editing the cal string message, press
Shift
>
ACAL (Up Arrow)
key.
Use the procedure below to use menus and softkeys to edit the cal string message and perform autocal
Step 1:
Press
Utility
on the front panel.
Step 2:
Press the
Test/Admin
softkey:
Step 3:
Press the
Calibrate
softkey:
Step 4:
(Optional step) Press the
Edit Cal String
softkey and enter the information for the cal string mes-
sage. Common messages include the last calibration date, calibration due date, or calibration department
contact information. When finished, press the
Done
softkey
Step 5:
Press the
Perform Autocal
softkey:
Keysight True
volt
Series Operating and Service Guide
139
Features and Functions
Содержание 34460A
Страница 1: ...Operating and Service Guide Keysight Truevolt Series Digital Multimeters ...
Страница 183: ...Measurement Tutorial 184 Keysight Truevolt Series Operating and Service Guide ...
Страница 385: ...STATus Subsystem SCPI Programming Reference 386 Keysight Truevolt Series Operating and Service Guide ...
Страница 541: ...Next Procedure Calibration Adjustment Procedures 542 Keysight Truevolt Series Operating and Service Guide ...
Страница 556: ...Next Procedure Keysight Truevolt Series Operating and Service Guide 557 Calibration Adjustment Procedures ...