Capacitance
This section describes how to configure capacitance measurements from the front panel.
Step 1:
Configure the test leads as shown.
Step 2:
Press
on the front panel.
Step 3:
To null–out the test lead capacitance:
l
Disconnect the + and - test leads probe end from the test circuit, and leave open.
l
Press
Null
. The DMM will now subtract this null value from capacitance measurements.
Step 4:
Press
Range
to select a range for the measurement. You can also use the
[+]
,
[-]
, and
[Range]
keys on the front panel to select the range. (
Auto
(autorange) automatically selects the range for the meas-
urement based on the input. Autoranging is convenient, but it results in slower measurements than using
a manual range. Autoranging goes down a range at less than 10% of range and up a range at greater than
120% of range. For capacitance measurements only, when autorange is off, the instrument does not
report an overload for readings greater than 120% of range. Overload only occurs when the algorithm
times out because the applied capacitance is too large for the algorithm to measure. If you apply a DC
voltage or a short to the input terminals in capacitance measurement mode, the instrument reports an
overload.
Features and Functions
76
Keysight True
volt
Series Operating and Service Guide
Содержание 34460A
Страница 1: ...Operating and Service Guide Keysight Truevolt Series Digital Multimeters ...
Страница 183: ...Measurement Tutorial 184 Keysight Truevolt Series Operating and Service Guide ...
Страница 385: ...STATus Subsystem SCPI Programming Reference 386 Keysight Truevolt Series Operating and Service Guide ...
Страница 541: ...Next Procedure Calibration Adjustment Procedures 542 Keysight Truevolt Series Operating and Service Guide ...
Страница 556: ...Next Procedure Keysight Truevolt Series Operating and Service Guide 557 Calibration Adjustment Procedures ...