[SENSe:]VOLTage[:DC]:RATio:SECondary {"OFF"|"CALCulate:DATA"|"SENSe:DATA"}
[SENSe:]VOLTage[:DC]:RATio:SECondary?
Selects the secondary measurement function for DC ratio measurements.
Parameter
Typical Return
{"OFF"|"CALCulate:DATA"|"SENSe:DATA"}, default
"OFF"
"CALC:DATA"
Configure the instrument for DC voltage ratio measurements using the 100 V range with 1 mV resolution.
Select "SENSe:DATA" as the secondary measurement. Make and retrieve measurements.
CONF:VOLT:DC:RAT 100,0.001
VOLT:RAT:SEC "SENS:DATA"
READ?;DATA2?
Typical Response: +1.0500;+2.1000,+1.9900
l
"CALCulate:DATA" - (Applies only to the 34465A and 34470A.) The measurement value before any
math operations are done (including NULL).
l
“SENSe:DATA” - The DC signal voltage and the DC reference voltage measurements.
l
This parameter is set to its default value after a Factory Reset (
) or Instrument Preset
(
).
[SENSe:]VOLTage[:DC]:RESolution {<
resolution
>|MIN|MAX|DEF}
[SENSe:]VOLTage[:DC]:RESolution? [{MIN|MAX|DEF}]
Selects the measurement resolution for DC voltage and ratio measurements. Specify the resolution in the
same units as the selected measurement function, not in number of digits.
Parameter
Typical Return
<
resolution
>: See
or
The default is
equivalent to 10 PLC. Specify <
resolution
> in the measurement's units (V, A, Hz, Ω,
and so on).
+3.00000000E-05
Configure DC voltage measurements with a 3 µV resolution. Then make and read one measurement:
CONF:VOLT:DC 1
VOLT:DC:RES 3E-6
READ?
l
You can specify MIN (best resolution) or MAX (worst resolution) instead of <
resolution
>.
l
To achieve normal mode (line frequency noise) rejection, use a resolution that corresponds to an integ-
ration time that is an integral number of power line cycles.
l
This parameter is set to its default value after a Factory Reset (
) or Instrument Preset
(
).
Keysight True
volt
Series Operating and Service Guide
383
SCPI Programming Reference
Содержание 34460A
Страница 1: ...Operating and Service Guide Keysight Truevolt Series Digital Multimeters ...
Страница 183: ...Measurement Tutorial 184 Keysight Truevolt Series Operating and Service Guide ...
Страница 385: ...STATus Subsystem SCPI Programming Reference 386 Keysight Truevolt Series Operating and Service Guide ...
Страница 541: ...Next Procedure Calibration Adjustment Procedures 542 Keysight Truevolt Series Operating and Service Guide ...
Страница 556: ...Next Procedure Keysight Truevolt Series Operating and Service Guide 557 Calibration Adjustment Procedures ...